FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 2052469 · Received March 16, 2011

Report

Report Number
3002158293-2011-00314
Event Type
Malfunction
Date Received
March 16, 2011
Date of Event
November 30, 2010
Report Date
March 14, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. UPON EVALUATION, THE CABLE RUNNING FROM THE DISTRIBUTION NODE TO ECG WAS CUT. THE ROOT CAUSE FOR THE DAMAGED CABLE CANNOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE.

Description of Event or Problem · 1

A REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, ELECTRODE BELT (B)(4) HAD A DAMAGED CABLE. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY PROBLEMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 3000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA