FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE SPORT
MDR report key: 2052408
·
Received March 16, 2011
Report
- Report Number
- 1824206-2011-01647
- Event Type
- Malfunction
- Date Received
- March 16, 2011
- Date of Event
- February 15, 2011
- Report Date
- February 15, 2011
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
REPAIRS HAVE NOT BEEN COMPLETED.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE BED WAS ALARMING. THE BED WAS INSPECTED AND THEY FOUND THE LEFT HEAD SIDERAIL DETECTION SWITCH CABLE AND GROUND STRAP WERE CUT IN TWO. HE BELIEVES THESE TWO PARTS GOT CAUGHT ON THE FOOT HI/LOW BEARING BLOCK MOUNTING BOLT AND NUT AND WHEN THE HI/LOW FUNCTION WAS RAISED IT CAUSED THE SWITCH CABLE AND GROUND STRAP TO BE CUT IN TWO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE SPORT | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |