FDA Adverse Event Malfunction Summary report: N

TOTALCARE SPORT

MDR report key: 2052408 · Received March 16, 2011

Report

Report Number
1824206-2011-01647
Event Type
Malfunction
Date Received
March 16, 2011
Date of Event
February 15, 2011
Report Date
February 15, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REPAIRS HAVE NOT BEEN COMPLETED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BED WAS ALARMING. THE BED WAS INSPECTED AND THEY FOUND THE LEFT HEAD SIDERAIL DETECTION SWITCH CABLE AND GROUND STRAP WERE CUT IN TWO. HE BELIEVES THESE TWO PARTS GOT CAUGHT ON THE FOOT HI/LOW BEARING BLOCK MOUNTING BOLT AND NUT AND WHEN THE HI/LOW FUNCTION WAS RAISED IT CAUSED THE SWITCH CABLE AND GROUND STRAP TO BE CUT IN TWO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE SPORT AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1