FDA Adverse Event Malfunction Summary report: N

AUTOMATE 800 SAMPLE PROCESSING SYSTEM

MDR report key: 2052397 · Received April 12, 2011

Report

Report Number
2050012-2011-01070
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
March 14, 2011
Report Date
March 14, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
CLASS1EXEMPT
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AFTER THE EVENT, THE CUSTOMER DISCONNECTED THE UPS AND IS RUNNING THE AUTOMATE FROM HOUSE CURRENT. THE UPS UNIT IS INSTALLED BY BCI, BUT IS NOT A BCI PRODUCT; THE UPS UNIT IS A POWERVAR PRODUCT. BCI DOES NOT SERVICE THE UPS UNITS. BCI SERVICE NOTIFIED POWERVAR AND REQUESTED TO VISIT THE CUSTOMER AND SERVICE OR REPLACE THE UPS UNITS.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THE UNINTERRUPTED POWER SUPPLY (UPS) CONNECTED TO THE AUTOMATE 800 SYSTEM FAILED CAUSING SMOKE AND FLAMES COMING FROM THE POWER SUPPLY, WHICH LED TO A CALL TO THE FIRE DEPARTMENT. THERE WAS NO OPERATOR EXPOSURE TO SMOKE REPORTED AND NO DAMAGE TO THE AUTOMATE SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATE 800 SAMPLE PROCESSING SYSTEM SAMPLE PROCESSING SYSTEM JJE BECKMAN COULTER INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1