COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH
Report
- Report Number
- 6000001-2011-02806
- Event Type
- Malfunction
- Date Received
- April 12, 2011
- Date of Event
- August 6, 2010
- Report Date
- August 24, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Removal / Correction Number
- 6000001-12/1/-08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY ALARM WAS CONFIRMED. THE ASSIGNABLE CAUSE WAS DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS HAVE BEEN REPLACED. ADDITIONAL: SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
THE FIELD SERVICE TECHNICIAN FROM BAXTER (B)(4) SERVICED A COLLEAGUE INFUSION PUMP ONSITE FOR DAMAGED BATTERY ALARM. THE PROCESS STEP IS UNKNOWN. DURING BAXTER'S REVIEW OF THE EVENT HISTORY, IT WAS DISCOVERED THAT A REPLACE BATTERY SET ALARM OCCURRED DURING INFUSION, WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.08.92.