MAXIMO DR
Report
- Report Number
- 6000144-2011-02040
- Event Type
- Death
- Date Received
- April 12, 2011
- Date of Event
- November 20, 2005
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S37
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. THE DEVICE(S) ASSOCIATED WITH THIS ADVERSE OUTCOME WAS/WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
EVALUATION SUMMARY: (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION HAD A COSMETIC DEPRESSION. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT ALL CONDUCTORS DISTORTED AND THERE WAS APPARENT EXPLANT DAMAGE (B)(4): THE DEVICE WAS RETURNED AND ANALYZED AND FOUND TO HAVE NORMAL BATTERY DEPLETION. THE DEVICE(S) ASSOCIATED WITH THIS ADVERSE OUTCOME WAS/WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
AN IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) AND TWO LEADS WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. INFORMATION IDENTIFIED IN THE MANUFACTURE'S DATA BASE INDICATION THE PATIENT DIED APPROXIMATELY ONE MONTH POST THE IMPLANT OF THE ICD SYSTEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAXIMO DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | 7278 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Death |