FDA Adverse Event Malfunction Summary report: N

WAVEONE GOLD SMALL 6-FILE STER 25MM

MDR report key: 20522398 · Received October 24, 2024

Report

Report Number
8031010-2024-00128
Event Type
Malfunction
Date Received
October 24, 2024
Date of Event
August 19, 2024
Report Date
December 5, 2024
Manufacturer
MAILLEFER INSTRUMENTS HOLDING SARL
Product Code
EKS
UDI-DI
D716WGSMALL251
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AS A RESULT OF THIS MALFUNCTION, THE POTENTIAL FOR SURGICAL INTERVENTION EXISTS TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION AS EVIDENCED BY PREVIOUS REPORTED EVENTS WITH SIMILAR FILES. THIS EVENT, THEREFORE, IS REPORTABLE PER 21CFR PART 803. THE DEVICE WAS NOT RETURNED FOR EVALUATION. HOWEVER, THE LOT NUMBER WAS PROVIDED AND RETAINED-PRODUCT TESTING AND/OR DHR REVIEW ARE PLANNED. THE RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS SUMMARY: INVOLVED PRODUCT THAT BROKE DURING USE WAS NOT RETURNED AND CANNOT BE ANALYZED. MOREOVER, NO UNUSED FILE IS AVAILABLE FOR EVALUATION. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHR REVIEW (BATCH #1820873). ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND. POTENTIAL ROOT CAUSES MAY BE INCORRECT TECHNIQUE, OVERUSE (MUPLTIPLE REUSES MAY INCREASE RISKS OF BREAKAGE- NUMBER OF USE NOT FORMALLY KNOWN), EXCESSIVE WEAR, OR MATERIAL ISSUE (NO ANALYSIS OF THE BROKEN FRAGMENTS POSSIBLE). WE MOREOVER DRAW CUSTOMER ATTENTION ABOUT THE FACT THAT WAVEONE GOLD FILES ARE SINGLE USE ITEMS.

Description of Event or Problem · 0

IN THIS EVENT IT IS REPORTED THAT A WAVEONE GOLD SMALL 6-FILE STER 25MM BROKE DURING USE. THE BROKEN PART WAS NOT RETRIEVED AND WAS INCORPORATED INTO THE FILLING. NO INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1714235 WAVEONE GOLD SMALL 6-FILE STER 25MM FILE, PULP CANAL, ENDODONTIC EKS MAILLEFER INSTRUMENTS HOLDING SARL 1820873 D716WGSMALL251

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown