FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2052219 · Received April 12, 2011

Report

Report Number
1423500-2011-04397
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
November 24, 2010
Report Date
April 6, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER DISCONTINUED USE OF THE DEVICE AND RETURNED IT TO THE (B)(4) FACILITY. REVIEW OF THE DEVICE LOGS REVEALED A VOLUME OF FLUID MEETING INCREASED INTRAPERITONEAL VOLUME (IIPV) CRITERIA. THE DEVICE WAS DETERMINED TO MEET SPECIFICATIONS RELATIVE TO THE IIPV IN THE LOGS. THE ASSIGNABLE CAUSE OF THE IIPV WAS DETERMINED TO BE: INSUFFICIENT DRAIN-ONE OR MORE CYCLE ADVANCES TO NEXT FILL WHEN SLOW/ NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. REVIEW OF THE SERVICE DATES AND ACTIVITIES REVEALED THE PREVIOUS RETURN OF THE DEVICE WAS NOT FOR THE REPORTED PROBLEMS OF IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) THAT WAS IDENTIFIED IN PATIENT THERAPY LOG WITH UF VOLUME OF 1118 ML DURING CYCLE 4. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED. THIS EVENT MEETS OVERFILL CRITERIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 15 YR HOMECHOICE AUTOMATED PD SET WITH CASSETTE