ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2011-04662
- Event Type
- Injury
- Date Received
- April 12, 2011
- Date of Event
- February 21, 2011
- Report Date
- February 21, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE RIGHT VENTRICULAR LEAD WAS SURGICALLY ABANDONED AND REPLACED. THE DEVICE WAS REMOVED, REPLACED, AND RETURNED FOR ANALYSIS. AS THE ABANDONED LEAD WAS NOT RETURNED FOR ANALYSIS, THE CLINICAL OBSERVATIONS OF THE LEAD COULD NOT BE CONFIRMED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED AND AN AMENDED REPORT SUBMITTED SHOULD FURTHER INFORMATION BE PROVIDED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT IMPLANTED WITH THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR UNDERWENT A PULMONARY ABLATION DUE TO ATRIAL FIBRILLATION. DURING THE PROCEDURE EXTERNAL DEFIBRILLATION WAS REQUIRED. THE PATCHES OF THE EXTERNAL DEFIBRILLATOR HAD BEEN ATTCHED ON THE LEFT AND RIGHT SIDES OF THE PATIENT'S CHEST. AFTER THE ABLATION PROCEDURE THE DEVICE WAS INTERROGATED AND HIGHER THAN BEFORE STIMULATIONS THRESHOLDS AND PACING IMPEDANCES ON THE RIGHT VENTRICULAR DEFBIRLLATION LEAD AND LEFT VENTRICULAR LEAD WERE FOUND. IN ADDITION, A LOW SHOCK IMPEDANCE MEASUREMENT WAS NOTED DUE TO AN ALERT MESSAGE OF LOW SHOCK IMPEDANCE LESS THAN 20 OHMS NOTIFICATION. THIS WAS LIKELY A RESULT OF THE COMMANDED 31J SHOCK REQUIRED DURING THE ABLATION PROCEDURE. THIS WAS A COMMANDED SHOCK FOR CARDIOVERSION TO TERMINATE ATRIAL FIBRILLATION. THE DEVICE WAS INTERROGATED; THE ATRIAL LEAD REVEALED NORMAL VALUES AND PARAMETERS. THE RIGHT VENTRICULAR SENSING WAS NORMAL, HOWEVER, HIGH PACING IMPEDANCE AND HIGH THRESHOLDS WERE NOTED. THE SHOCK IMPEDANCE WAS NORMAL DURING INTERROGATION. THE LEFT VEDNTRICULAR IMPEDANCE WAS GREATER THAN 2,000 OHMS, AND THRESHOLDS WERE HIGH. ALL OTHER MEASUREMENTS WERE NORMAL. A REVISION PROCEDURE WAS PERFORMED; UPON REMOVAL OF THE DEVICE, A BLACK MARK WAS NOTED ON THE DEVICE. THE RIGHT VENTRICULAR LEAD HAD BEEN LYING ON THE BLACK MARK OF THE DEVICE AND THE LEAD INSULATION WAS DAMAGED. A SHORTED LEAD CONDITION WAS SUSPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0181 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Life Threatening| R |