FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL TIP 1ML

MDR report key: 20521626 · Received October 24, 2024

Report

Report Number
1911916-2024-00753
Event Type
Malfunction
Date Received
October 24, 2024
Date of Event
October 10, 2024
Report Date
October 25, 2024
Manufacturer
BECTON DICKINSON
Product Code
FMF
UDI-DI
30382903096535
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

PR (B)(4) FOLLOW UP FOR DEVICE EVALUATION. IT WAS REPORTED THE SYRINGE LEAKS FROM THE PLUNGER SIDE WHEN DRAWING UP MEDICATION. TO AID IN THE INVESTIGATION, ONE SAMPLE IN AN OPENED PACKAGING BLISTER WAS RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED. FIRST, WITH 10X MAGNIFICATION, AND THEN WITH 30X MAGNIFICATION. NO DEFECTS OR IMPERFECTIONS WERE OBSERVED. A FUNCTIONAL TEST WAS THEN PERFORMED BY DRAWING 50ML OF SALINE SOLUTION INTO THE SYRINGE. NO LEAKAGE OR ANY OTHER ISSUE WAS OBSERVED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309653, LOT 4180131. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED.

Description of Event or Problem · 0

MATERIAL #:309653; BATCH#: UNKNOWN. IT WAS REPORTED BY CUSTOMER THAT THE SYRINGE LEAKS FROM PLUNGER SIDE WHEN DRAWING UP MEDICATION.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1322924 SYRINGE 50ML LL TIP 1ML SYRINGE, PISTON FMF BECTON DICKINSON 4180131 30382903096535

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown