FDA Adverse Event Malfunction Summary report: N

HUDSON SINGLE-HEATED LIMB CIRCUIT ACCESSORY

MDR report key: 2052132 · Received March 1, 2011

Report

Report Number
3004365956-2011-00090
Event Type
Malfunction
Date Received
March 1, 2011
Date of Event
January 30, 2011
Report Date
February 8, 2011
Manufacturer
TELEFLEX MEDICAL
Product Code
BZO
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVAL: METHOD - MANUFACTURING PROCESS REVIEW. RESISTANCE MEASUREMENTS OF INSPIRATORY AND EXPIRATORY WIRES. RESULTS - VISUAL EXAMINATION CONFIRMS A MELT IN THE TUBING OF THE CIRCUIT. THE DAMAGE CONSIST OF BOTH COLLAPSED TUBING AND MELTING THROUGH THE HOLES IN THE TUBING. THE MANUFACTURING PROCESS REVIEW OF CATALOG #780-20 WAS REVIEWED AND NO ISSUES WERE DETECTED. THE RESISTANCE OF THE INSPIRATORY AND EXPIRATORY WIRES MEASURED CORRECTLY AT 13.3 OHMS. THIS IS WITHIN THE ALLOWABLE RANGE OF 12.8 TO 14.3 OHMS. THE TYPE OF MELT IS TYPICALLY DUE TO CUSTOMER MISUSE. THE CIRCUIT WAS MOST LIKELY COVERED BY A BLANKET.

Description of Event or Problem · 1

THE EVENT IS REPORTED AS: THE CIRCUIT MELTED WHILE IN USE ON A PATIENT. THE NURSE FOUND THE UPPER PART OF THE HIGH FLOW CIRCUIT MELTED AND STUCK TO THE BED LINEN. THE PATIENT'S OXYGEN SATURATION LEVEL DECREASED. THE PATIENT WAS GIVEN OXYGEN VIA BAG AND MASK AND THE PATIENT'S OXYGEN SATURATION INCREASED. IT IS REPORTED THAT THE PATIENT IS DOING FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HUDSON SINGLE-HEATED LIMB CIRCUIT ACCESSORY VENTILATOR CIRCUIT BZO TELEFLEX MEDICAL NA UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 UNK HUDSON CONCHA NEPTUNE: CATALOG# 425-00| (MDR# 3003898360-2011-00073)