HUDSON SINGLE-HEATED LIMB CIRCUIT ACCESSORY
Report
- Report Number
- 3004365956-2011-00090
- Event Type
- Malfunction
- Date Received
- March 1, 2011
- Date of Event
- January 30, 2011
- Report Date
- February 8, 2011
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- BZO
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVAL: METHOD - MANUFACTURING PROCESS REVIEW. RESISTANCE MEASUREMENTS OF INSPIRATORY AND EXPIRATORY WIRES. RESULTS - VISUAL EXAMINATION CONFIRMS A MELT IN THE TUBING OF THE CIRCUIT. THE DAMAGE CONSIST OF BOTH COLLAPSED TUBING AND MELTING THROUGH THE HOLES IN THE TUBING. THE MANUFACTURING PROCESS REVIEW OF CATALOG #780-20 WAS REVIEWED AND NO ISSUES WERE DETECTED. THE RESISTANCE OF THE INSPIRATORY AND EXPIRATORY WIRES MEASURED CORRECTLY AT 13.3 OHMS. THIS IS WITHIN THE ALLOWABLE RANGE OF 12.8 TO 14.3 OHMS. THE TYPE OF MELT IS TYPICALLY DUE TO CUSTOMER MISUSE. THE CIRCUIT WAS MOST LIKELY COVERED BY A BLANKET.
THE EVENT IS REPORTED AS: THE CIRCUIT MELTED WHILE IN USE ON A PATIENT. THE NURSE FOUND THE UPPER PART OF THE HIGH FLOW CIRCUIT MELTED AND STUCK TO THE BED LINEN. THE PATIENT'S OXYGEN SATURATION LEVEL DECREASED. THE PATIENT WAS GIVEN OXYGEN VIA BAG AND MASK AND THE PATIENT'S OXYGEN SATURATION INCREASED. IT IS REPORTED THAT THE PATIENT IS DOING FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HUDSON SINGLE-HEATED LIMB CIRCUIT ACCESSORY | VENTILATOR CIRCUIT | BZO | TELEFLEX MEDICAL | NA | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | HUDSON CONCHA NEPTUNE: CATALOG# 425-00| (MDR# 3003898360-2011-00073) |