VOLCANO S5/S5I SERIES INTRAVASCULAR IMAGING

MDR report key: 2052056 · Received March 2, 2011

Report

Report Number
2939520-2011-00006
Event Type
Malfunction
Date Received
March 2, 2011
Date of Event
February 2, 2011
Report Date
February 2, 2011
Manufacturer
VOLCANO CORPORATION
Product Code
IYO
PMA / PMN Number
K082229
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PC MODULE WAS RETURNED TO VOLCANO FOR EVALUATION. THE SIDE PANEL WAS REMOVED AND THE ALL COMPONENTS WERE INSPECTED UNDER A LIGHT MICROSCOPE. DISCOLORATION WAS OBSERVED UNDER THE CHIP FOLLOWING THE CAT5 CABLE CONNECTION. THE PINS ON THE CHIP WERE REMOVED TO FURTHER EXAMINE THE CHIP; IT WAS OBSERVED THAT THE WIRE COATINGS WERE BURNT. THIS CHIP IS A PASSIVE COMPONENT, WHICH SUGGESTS AN OVER-CURRENT WAS APPLIED; HOWEVER, WE COULD NOT CONCLUSIVELY DETERMINE THE ROOT CAUSE OF THIS FAILURE. A REPLACEMENT PC MODULE WAS INSTALLED AND THE SYSTEM IS FUNCTIONING AS INTENDED. NO FURTHER ACTION REQUIRED.

Description of Event or Problem · 1

DURING SYSTEM SERVICE BY A FIELD SERVICE ENGINEER (FSE) DUE TO A REPORTED IMAGE ISSUE, A PC MODULE WAS REPLACED. POST REPLACING THE PC MODULE, IT WAS NOTICED THAT THE EXTERNAL USB EXTENDER (REX) WAS DAMAGED, SPECIFICALLY, THE PORT CONNECTION ON THE REX WAS BROKEN. THIS WAS DUE TO THE BEDSIDE CONTROL CONSOLE CABLES BEING WRAPPED AROUND THE TABLE SIDE; THUS THE REX WAS REPLACED. POST REPLACING THE REX, THE SYSTEM WAS FUNCTIONAL FOR A FEW PROCEDURES AND THEN EXPERIENCED BEDSIDE CONTROL LOCK UP; AND IT WAS NOTICED THAT THE REX LIGHTS DID NOT ILLUMINATE. IN TURN THE REX WAS AGAIN TESTED BY THE FSE ON ANOTHER FUNCTIONAL SYSTEM AND IT DID ILLUMINATE; THEN HE REMOVED THE SIDE PANEL FROM THE PC MODULE AND DISCONNECTED THE INTERNAL CAT5 CABLE TO INSPECT THE CAT5 BULKHEAD CONNECTOR. AT THIS TIME THE FSE INDICATED HE COULD SMELL SOMETHING SIMILAR TO A HOT COMPONENT AND SMOKE WAS NOTED. THE DEVICE WAS REMOVED FROM SERVICE. THERE WAS NO PT HARM OR SERIOUS ADVERSE EVENT.