FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 2051968 · Received April 12, 2011

Report

Report Number
2649622-2011-05708
Event Type
Injury
Date Received
April 12, 2011
Date of Event
January 30, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT CAME TO THE EMERGENCY ROOM DUE TO POCKET STIMULATION. THE ATRIAL LEAD SHOWED LOW IMPEDANCE, IMPEDANCE WAS HIGHER IN UNIPOLAR THAN IN BIPOLAR, AND THE POLARITY HAD SWITCHED. THE LEAD IS STILL IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4068 ASKU

Patients

Seq Age Sex Outcome Treatment
1 55 YR Hospitalization (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD