FDA Adverse Event Malfunction Summary report: N

ATTAIN OTW

MDR report key: 2051966 · Received April 12, 2011

Report

Report Number
2649622-2011-05707
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
January 24, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S012
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED, THE PROXIMAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER INSULATION WAS BREACHED CUT, AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT AT IMPLANT ATTEMPT, TWO DIFFERENT LEADS WERE OBSERVED TO BE FRACTURED. THE LEADS WERE NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4194 ASKU

Patients

Seq Age Sex Outcome Treatment
1 53 YR Other