FDA Adverse Event Death Summary report: N

QDOT MICRO

MDR report key: 20519220 · Received October 23, 2024

Report

Report Number
2029046-2024-03462
Event Type
Death
Date Received
October 23, 2024
Date of Event
September 9, 2024
Manufacturer
BIOSENSE WEBSTER INC
Product Code
OAE
UDI-DI
10846835016758
PMA / PMN Number
P210027
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

DEVICE INVESTIGATION DETAILS: AN ANALYSIS OF THE PRODUCT COULD NOT BE PERFORMED SINCE A PHYSICAL SAMPLE WAS NOT RECEIVED FOR EVALUATION. HOWEVER, IF THE PRODUCT IS RECEIVED AT A LATER DATE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE NUMBER LOT 31378343L AND NO INTERNAL ACTION RELATED TO THE COMPLAINT WAS FOUND DURING THE REVIEW. AS PART OF OUR COMPANY QUALITY SYSTEM PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND DISTRIBUTED TO APPROVED SPECIFICATIONS. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY BIOSENSE WEBSTER INC., OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, BIOSENSE WEBSTER INC., OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MANUFACTURER'S REF. # (B)(4).

Additional Manufacturer Narrative · 0

ON 24-OCT-2024, ADDITIONAL INFORMATION WAS RECEIVED CLARIFYING A STATEMENT PREVIOUSLY REPORTED IN SECTION B5. EVENT DESCRIPTION OF THE 3500A INITIAL MEDWATCH. WHAT THEY MEANT WAS, "IT IS CURRENTLY UNKNOWN WHAT FACTORS, INCLUDING ANATOMICAL STRUCTURES OF THE VERTEBRAE, ETC., MAY HAVE CONTRIBUTED TO THE EVENT.¿ IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REF. # (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT A PULMONARY VEIN ISOLATION (PVI) AND POSTERIOR WALL ABLATION PROCEDURE WITH A QDOT MICRO AND A NGEN GENERATOR AND THE PATIENT EXPERIENCED ATRIO-ESOPHAGEAL FISTULA / PERFORATION THAT LED TO SEPSIS AND CEREBRAL INFARCTION. THE PATIENT DIED. ON (B)(6), ABLATION WAS PERFORMED AT MIE HEART CENTER (PVI + POSTERIOR WALL ISOLATION), NO ISSUE WAS FOUND DURING THE PROCEDURE. ON (B)(6), THE PATIENT WAS TRANSFERRED TO ((B)(6) HOSPITAL DUE TO UNSTABLE VITAL SIGNS. ON (B)(6), THE PATIENT WAS TRANSFERRED TO (B)(6) HOSPITAL DUE TO IMPAIRED CONSCIOUSNESS. ON COMPUTED TOMOGRAPHY (CT), ESOPHAGEAL PERFORATION AND MEDIASTINAL EMPHYSEMA WERE CONFIRMED, SO THE PATIENT WAS TRANSFERRED TO (B)(6) HOSPITAL. ON (B)(6) THE PATIENT'S CONDITION WORSENED AND HE DIED OF SEPSIS AND CEREBRAL INFARCTION. THE PHYSICIAN'S OPINION ON THE RELATIONSHIP BETWEEN THE EVENT AND THE PRODUCT IS THAT THERE WERE NO DEFECTIVE PRODUCTS AND THAT IT IS POSSIBLE THAT THE FACT THAT THE AREA NEAR THE ESOPHAGUS WAS ABLATED TWICE WITH 90W WITH THE QDOT MICRO CATHETER HAD AN EFFECT, BUT THE DIRECT CAUSE IS CURRENTLY UNKNOWN. THE PHYSICIAN REPORTED THAT THE EFFECT ON THE ESOPHAGUS WAS RATHER CONSIDERED AND VHPSD (HIGH POWER SHORT DURATION) WAS FELT TO BE SAFE. THEREFORE, SHORT TERM POWER SUPPLY + SKIP POWER SUPPLY (TO PREVENT CONTINUOUS POWER SUPPLY AROUND ESOPHAGUS) WAS APPLIED WHILE CONTACT FORCE (CF) WAS SUPPRESSED. THEREFORE, IT WAS NOT BECAUSE OF THE HIGH OUTPUT POWER OR ANYTHING LIKE THAT. AT THE MOMENT, IT IS UNCLEAR "WHAT AFFECTED THE VERTEBRAL BODY", INCLUDING THE ANATOMY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1375190 QDOT MICRO CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION OAE BIOSENSE WEBSTER INC 31378343L 10846835016758

Patients

Seq Age Sex Outcome Treatment
1 86 YR Female Death 10439011| 7FR,DEF,15MM DIA., CCW,LASSO| NGEN| NGEN PUMP, JAPAN CONFIGURATION| OCTA,STD,48P,2-2-2-2-2,D-CURVE| UNK_CARTO 3