FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2051920 · Received April 12, 2011

Report

Report Number
2649622-2011-05690
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
January 21, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND THE HELIX HAD DISENGAGED FROM THE HELICAL CHANNEL. IT WAS ALSO NOTED THAT THE INNER TUBING WAS KINKED/BUCKLED AND THERE WAS BLOOD IN/ON THE HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE LEAD REQUIRED RE-POSITIONING SEVERAL TIMES. DURING ONE OF THE RE-POSITIONINGS, THE HELIX WOULD NOT EXTEND. THE LEAD WAS EXTRACTED AND TESTED ON THE TABLE, BUT THE HELIX STILL WOULD NOT EXTEND. THE LEAD WAS ATTEMPTED AND NOT USED AND A NEW LEAD WAS IMPLANTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other