FDA Adverse Event Malfunction Summary report: N

ENRHYTHM DR

MDR report key: 2051918 · Received April 12, 2011

Report

Report Number
6000094-2011-00511
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
September 22, 2009
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S38
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS CLOSE TO ELECTIVE REPLACEMENT INDICATOR (ERI) A DAY AFTER IMPLANT. INTERROGATION SHOWED THE DEVICE HAD OUTPUT SINCE APPROXIMATELY A YEAR AND A HALF PRIOR TO IMPLANT. THE DEVICE IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENRHYTHM DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. P1501DR ASKU

Patients

Seq Age Sex Outcome Treatment
1 75 Other