FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 2051899 · Received April 12, 2011

Report

Report Number
2124215-2011-03371
Event Type
Injury
Date Received
April 12, 2011
Date of Event
February 23, 2011
Report Date
April 26, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL AVAILABLE INFORMATION INDICATES THAT THE LEAD REMAINS IN SERVICE. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT A LOW SHOCK IMPEDANCE MEASUREMENT WAS DETECTED ON THIS DEVICE SYSTEM. THE PATIENT WAS BROUGHT INTO THE CLINIC FOR FURTHER EVALUATION. OUT OF RANGE IMPEDANCE MEASUREMENTS WERE UNABLE TO BE RECREATED. TECHNICAL SERVICES DISCUSSED TESTING RECOMMENDATIONS. TO DATE, NO ADVERSE PATIENT EFFECTS WERE REPORTED WITH THIS CLINICAL OBSERVATION.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS REPORTED THAT A NON-INVASIVE PROGRAMMED STIMULATION (NIPS) PROCEDURE WAS PERFORMED. SHOCKS DELIVERED AT 1.1 JOULES AND 41 JOULES EXHIBITED IMPEDANCE MEASUREMENTS WITHIN ACCEPTABLE LIMITS. A DEFIBRILLATION THRESHOLD TEST WAS ALSO SUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0185