FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2051865 · Received April 12, 2011

Report

Report Number
2649622-2011-05677
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
January 19, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND THE DISTAL CONDUCTOR WAS DISTORTED. IT WAS ALSO NOTED THAT THE DISTAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THE INNER TUBING WAS KINKED/BUCKLED, THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND SLEEVE HEAD, THE LEAD WAS STRETCHED, AND THE LEAD WAS DAMAGED AT IMPLANT. THE ANALYST ALSO NOTED THAT THE HELIX CANNOT EXTEND AND RETRACT DUE TO DISTAL CONDUCTOR DISTORTION WITHIN THE CONNECTOR.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT THERE WAS DIFFICULTY POSITIONING THE LEAD, AND THE HELIX BECAME STUCK AFTER THE THIRD POSITIONING ATTEMPT. THE LEAD WAS NOT IMPLANTED, AND ANOTHER LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 84 YR Other