FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 2051846 · Received April 12, 2011

Report

Report Number
2124215-2011-03426
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
February 23, 2011
Report Date
February 23, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR (B)(4) LABORATORY, ICD MEMORY WAS REVIEWED. WE CONFIRMED THAT THE DEVICE DECLARED ERI WHILE IMPLANTED AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT OF 18.9 SECONDS. TO DETERMINE IF THE DEVICE'S RATE OF BATTERY DEPLETION WAS NORMAL, OUR TECHNICIANS COMPARED THE OBSERVED RATE OF BATTERY USAGE TO THE EXPECTED RATE OF BATTERY USAGE. THE RESULTS SHOWED THAT THE MONITORING VOLTAGE WAS NORMAL BASED ON THERAPY USE TO DATE AND PROGRAMMED SETTINGS. LABORATORY TECHNICIANS AND ENGINEERS CONCLUDED THAT THIS DEVICE DID NOT EXPERIENCE PREMATURE BATTERY DEPLETION. RATHER, AN ERI CHARGE TIME REPLACEMENT INDICATOR WAS DECLARED DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED. THERE WAS NO ALLEGATION FROM THE FIELD REGARDING THE FUNCTION OF THE DEVICE. THIS EVENT WAS CREATED DUE TO THE INITIAL TESTING PERFORMED BY OUR POST MARKET QUALITY ASSURANCE LABORATORY. INITIAL TESTING NOTED A POSSIBLE PERFORMANCE ISSUE CONCERNING BATTERY LONGEVITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T167

Patients

Seq Age Sex Outcome Treatment
1 63 YR (B)(4)| 1861| (B)(4)| T167| (B)(4)