PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 2051843
·
Received March 15, 2011
Report
- Report Number
- 2032227-2011-00651
- Event Type
- Malfunction
- Date Received
- March 15, 2011
- Date of Event
- February 28, 2011
- Report Date
- February 28, 2011
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARMED WITH MOTOR ERROR DURING THE DISPLACEMENT TEST DUE TO AN OUT OF PHASE ENCODER SIGNAL.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A MOTOR ERROR ALARM DURING A BOLUS. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC MINIMED | MMT-523RNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR |