FDA Adverse Event Malfunction Summary report: N

LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 2051783 · Received March 15, 2011

Report

Report Number
3015876-2011-00238
Event Type
Malfunction
Date Received
March 15, 2011
Date of Event
February 14, 2011
Report Date
February 14, 2011
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K991910
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE THERAPY CONNECTOR ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REMOVED THERAPY CONNECTOR ASSEMBLY AT THE FAILURE ANALYSIS CENTER AND DETERMINED THE CAUSE OF THE FAILURE TO BE BROKEN OFF CONTACT TABS IN SOCKET SEVEN AND THREE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD INTERMITTENTLY FAIL TO DETECT THE THERAPY CABLE CONNECTION WHEN IT WAS MOVED OR STRETCHED. THERE WAS NO PT USE ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES MKJ PHYSIO-CONTROL, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA