FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES
MDR report key: 2051783
·
Received March 15, 2011
Report
- Report Number
- 3015876-2011-00238
- Event Type
- Malfunction
- Date Received
- March 15, 2011
- Date of Event
- February 14, 2011
- Report Date
- February 14, 2011
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- MKJ
- PMA / PMN Number
- K991910
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE THERAPY CONNECTOR ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE. PHYSIO FURTHER EVALUATED THE REMOVED THERAPY CONNECTOR ASSEMBLY AT THE FAILURE ANALYSIS CENTER AND DETERMINED THE CAUSE OF THE FAILURE TO BE BROKEN OFF CONTACT TABS IN SOCKET SEVEN AND THREE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD INTERMITTENTLY FAIL TO DETECT THE THERAPY CABLE CONNECTION WHEN IT WAS MOVED OR STRETCHED. THERE WAS NO PT USE ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES | MKJ | PHYSIO-CONTROL, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |