FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2051740 · Received April 12, 2011

Report

Report Number
2649622-2011-05657
Event Type
Injury
Date Received
April 12, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S17
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DEFIBRILLATION COIL WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), THE DEFIBRILLATION CONDUCTOR WAS FRACTURED (OVERSTRESS), THE PROXIMAL CONDUCTOR WAS FRACTURED (OVERSTRESS), OUTER TUBING OVERLAY COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER INSULATION WAS TORN, EXPOSED DEFIB COIL WHITE SUBSTANCE, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS APPARENT EXPLORE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. HIGH RESISTANCE/IMPEDANCE WAS NOTED WITH ONE PATIENT ALERT FOR OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE ON (B)(6) 2010 02:15:03. THE WEEKLY PACE LEAD IMPEDANCE TREND DATA SHOWS MIN AND MAX RV PACE FROM 704 TO 3296 OHMS, BETWEEN (B)(6) 2010 AND (B)(6) 2011.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD HIGH IMPEDANCE WHICH INCREASED STEADILY UNTIL IT REACHED AN OUT OF RANGE LEVEL. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6944 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB