FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 2051726
·
Received April 12, 2011
Report
- Report Number
- 2649622-2011-05642
- Event Type
- Injury
- Date Received
- April 12, 2011
- Date of Event
- January 27, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DISTAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED. THERE WERE NO ANOMALIES FOUND. THERE WAS APPARENT EXPLANT DAMAGE. THE PROXIMAL CONDUCTOR WAS DISTORTED AND HAD BLOOD/BODY FLUID (NOT OBSTRUCTED). THE OUTER INSULATION HAD A BREACHED CUT. THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD DISLODGED POST-IMPLANT. DURING THE REPOSITIONING ATTEMPT THE LEAD KINKED DUE TO DIFFICULT PATIENT ANATOMY. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | Hospitalization| R | 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR |