FDA Adverse Event
Injury
Summary report: N
ENRHYTHM DR
MDR report key: 2051714
·
Received April 12, 2011
Report
- Report Number
- 6000144-2011-01542
- Event Type
- Injury
- Date Received
- April 12, 2011
- Date of Event
- December 21, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- DXY
- PMA / PMN Number
- P980035/S38
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND REVEALED A POWER ON RESET. A CRITICAL RAM PARITY ERROR OCCURED ON (B)(6) 2010. THE POR TYPE IS CONSIDERED LOW SEVERITY AND THE DEVICE SHOULD BE ABLE TO FULLY RECOVER AFTER RESET.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE EXPERIENCED A POWER ON RESET. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENRHYTHM DR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC MED REL, INC. | P1501DR | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 88 YR | Required Intervention | 5076 (X2) IMPLANTABLE PACING LEAD |