FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2051664 · Received April 12, 2011

Report

Report Number
2649622-2011-05624
Event Type
Injury
Date Received
April 12, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD POCKET STIMULATION. THE RIGHT VENTRICULAR LEAD HAD HIGH THRESHOLD AND INTERMITTENT CAPTURE. THE LEAD WAS CAPPED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 90 YR Hospitalization| R SESR01 IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD