SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-05603
- Event Type
- Injury
- Date Received
- April 12, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE PARTIAL LEAD IN SEGMENTS WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT ALL CONDUCTORS WERE DISTORTED, THAT SEVERAL CONDUCTORS HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THE OUTER TUBING OVERLAY HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE MEDIAL SECTION OF THE LEAD'S INSULATION WAS TORN AND WAS BREACHED CUT, THE OUTER OVERLAY TUBING HAD A WHITE SUBSTANCE, THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND SLEEVEHEAD, THE LEAD WAS STRETCHED, AND THERE WAS APPARENT EXPLANT DAMAGE.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD OVERSENSING AND COULD NOT BE REPOSITIONED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Hospitalization| R | 4574 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD |