FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2051602 · Received April 12, 2011

Report

Report Number
2649622-2011-05603
Event Type
Injury
Date Received
April 12, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE PARTIAL LEAD IN SEGMENTS WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT ALL CONDUCTORS WERE DISTORTED, THAT SEVERAL CONDUCTORS HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THE OUTER TUBING OVERLAY HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE MEDIAL SECTION OF THE LEAD'S INSULATION WAS TORN AND WAS BREACHED CUT, THE OUTER OVERLAY TUBING HAD A WHITE SUBSTANCE, THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND SLEEVEHEAD, THE LEAD WAS STRETCHED, AND THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD OVERSENSING AND COULD NOT BE REPOSITIONED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 56 YR Hospitalization| R 4574 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD