ATTAIN OTW
Report
- Report Number
- 2649622-2011-05592
- Event Type
- Malfunction
- Date Received
- April 12, 2011
- Date of Event
- January 27, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P010015/S012
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT AND THERE WAS APPARENT EXPLANT DAMAGE. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THERE WAS BLOOD/BODY FLUID ON THE OUTER TUBING OVERLAY AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.
IT WAS REPORTED THAT DURING IMPLANT, THE LEAD DISLODGED WHILE REMOVING THE LEFT VENTRICULAR SHEATH. A SECOND LEAD WAS ATTEMPTED BUT WOULD NOT WORK WITH THE PATIENT'S ANATOMY. A THIRD LEAD WAS ATTEMPTED AND IMPLANTED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN OTW | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4194 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 91 YR | Other | 4058M IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD |