CAPSUREFIX NOVUS

MDR report key: 2051562 · Received April 12, 2011

Report

Report Number
2649622-2011-05586
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
January 24, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THERE WAS BLOOD IN/ON THE HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE HELIX WOULD NOT DEPLOY AND THE PHYSICIAN COULD NOT GET LEAD TO CAPTURE ON THE RIGHT VENTRICULAR APEX. IT IS SUSPECTED THAT THERE MAY BE TISSUE IN THE HELIX OR THE HELIX MALFUNCTIONED. THE LEAD WAS EXPLANTED AND A NEW LEAD USED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 82 YR Other