FDA Adverse Event Malfunction Summary report: N

REAMER HANDLE OFFSET

MDR report key: 20515508 · Received October 23, 2024

Report

Report Number
3005985723-2024-00211
Event Type
Malfunction
Date Received
October 23, 2024
Date of Event
October 1, 2024
Report Date
March 6, 2026
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486032142
PMA / PMN Number
K170593
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

AS PART OF THE NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

AN EVENT REGARDING MISSING COMPONENT INVOLVING A MAKO REAMER HANDLE WAS REPORTED. THE EVENT WAS NOT CONFIRMED BECAUSE THE PRODUCT WAS NOT AVAILABLE FOR INSPECTION. METHOD & RESULTS: -PRODUCT EVALUATION AND RESULTS: MATERIAL ANALYSIS, VISUAL, FUNCTIONAL AND DIMENSIONAL INSPECTIONS COULD NOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED. -CLINICIAN REVIEW: NO MEDICAL RECORDS WERE RECEIVED FOR REVIEW WITH A CLINICAL CONSULTANT. -PRODUCT HISTORY REVIEW: A PRODUCT HISTORY REVIEW IS NOT REQUIRED, AS NO MANUFACTURING SPECIFIC ISSUE HAS BEEN ALLEGED. AT STRYKER JOINT REPLACEMENT ALL DEVICES/PRODUCT ARE MANUFACTURED THROUGH VALIDATED EQUIPMENT AND INSPECTED BY TRAINED REPRESENTATIVES THROUGH STRYKER¿S QUALITY MANAGEMENT SYSTEM. -COMPLAINT HISTORY REVIEW: A COMPLAINT HISTORY REVIEW AND TREND DETECTION IS NOT REQUIRED AS THIS IS A PM. MONITORING WILL BE CONTINUED UNDER THE CURRENT QUARTERLY TREND REVIEW. CONCLUSIONS: NO PATIENT WAS INVOLVED AND NO ACTUAL OR POTENTIAL PATIENT HARM EXISTED FOR THE ALLEGED EVENT. THE ALLEGED FAILURE MODE WAS NOT CONFIRMED BECAUSE THE PRODUCT WAS NOT AVAILABLE FOR INSPECTION. NO ADDITIONAL INVESTIGATION OR SPECIFIC ACTIONS ARE REQUIRED. IF ADDITIONAL INFORMATION IS RECEIVED THEN THE COMPLAINT WILL BE REOPENED.

Description of Event or Problem · 0

STERILE PROCESSING NOTICED SCREW MISSING FROM OFFSET REAMER DURING CLEANING/STERILIZATION. THA 4.1 APPLICATION, DISCOVERED DURING STERILE PROCESSING, NO PT UNDER ANESTHESIA, NO DELAY. CASE TYPE / APPLICATION: THA 4.1 - (IF APPLICABLE) WILL DEVICE BE DECONTAMINATED? DEVICE(S) DO NOT NEED TO BE DECONTAMINATED (NO PATIENT CONTACT/ BIOBURDEN).

Description of Event or Problem · 0

STERILE PROCESSING NOTICED SCREW MISSING FROM OFFSET REAMER DURING CLEANING/STERILIZATION. THA 4.1 APPLICATION, DISCOVERED DURING STERILE PROCESSING, NO PT UNDER ANESTHESIA, NO DELAY. CASE TYPE / APPLICATION: THA 4.1 - (IF APPLICABLE) WILL DEVICE BE DECONTAMINATED? DEVICE(S) DO NOT NEED TO BE DECONTAMINATED (NO PATIENT CONTACT/ BIOBURDEN).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
295570 REAMER HANDLE OFFSET ORTHOPEDIC STEREOTAXIC INSTRUMENT OLO MAKO SURGICAL CORP. 5561424 00848486032142

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other