1.5T SIGNA HDX ECHOSPEED
Report
- Report Number
- 2183553-2011-00014
- Event Type
- Injury
- Date Received
- April 7, 2011
- Date of Event
- February 14, 2011
- Report Date
- April 7, 2011
- Manufacturer
- GE MEDICAL SYSTEMS, LLC
- Product Code
- LNH
- PMA / PMN Number
- K052293
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A GEHC LOCAL FIELD TEAM WAS DISPATCHED TO THE CUSTOMER SITE TO OBTAIN SCAN SPECIFIC INFORMATION AND INVESTIGATE THE ENVIRONMENTAL CONDITIONS. THE INVESTIGATION FOCUSED ON FOUR MAIN AREAS: ENVIRONMENTAL: (INCLUDING COOLING EQUIPMENT, PT AIR COOLING PLUS THE MR SCANNER ERROR LOG). SURFACE COILS/ACCESSORIES: (SURFACE COIL/ECG CHECKS). SYSTEM/ IMAGE QUALITY: (SYSTEMS LEVEL TESTING TO CHECK FOR SYSTEM FAILURES). PT MONITORING: (PT ALARM & RF SAFETY MONITOR CHECKS). VISUAL INSPECTION AND THE TEST RESULTS FOR THE MRI SYSTEM SHOW NO ISSUES THAT CAUSED OR CONTRIBUTED TO THE BURN. THE SYSTEM WAS DETERMINED TO BE FUNCTIONING WITHIN SPECIFICATIONS. THE ROOT CAUSE OF THE INJURY COULD NOT BE DETERMINED.
IT WAS REPORTED THAT A PT SUSTAINED A SECOND DEGREE BURN ON THE MIDDLE OF HER BACK TO THE RIGHT OF THE MIDLINE DURING A MAGNETIC RESONANCE (MR) SCAN. THE PT WAS BEING SCANNED WITH THE 8 CHANNEL CTL COIL FOR A LUMBAR SPINE SCAN. THE PT WAS POSITIONED HEAD FIRST AND SUPINE, AND THE PT'S ARMS WERE POSITIONED OVER HER HEAD. THE PT REPORTEDLY FELT REDNESS AND DISCOMFORT, BUT SHE DID NOT INDICATE THIS DIRECTLY AFTER HER MR SCAN. SHE CONTACTED THE SITE 10 DAYS AFTER THE SCAN. THE HOSPITAL STATED THAT THE PT WAS PROPERLY ISOLATED FROM THE BORE. THE COMBINED SIZE OF THE BLISTERS WAS 2 X 3 CM. THE PULSE SEQUENCES USED WERE NOT PROVIDED. THE PT DID NOT SEEK ANY MEDICAL TREATMENT FOR THIS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 1.5T SIGNA HDX ECHOSPEED | MAGNETIC RESONANCE DIAGNOSTIC SYSTEM | LNH | GE MEDICAL SYSTEMS, LLC | 5308573 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Other |