FDA Adverse Event Injury Summary report: N

1.5T SIGNA HDX ECHOSPEED

MDR report key: 2051549 · Received April 7, 2011

Report

Report Number
2183553-2011-00014
Event Type
Injury
Date Received
April 7, 2011
Date of Event
February 14, 2011
Report Date
April 7, 2011
Manufacturer
GE MEDICAL SYSTEMS, LLC
Product Code
LNH
PMA / PMN Number
K052293
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GEHC LOCAL FIELD TEAM WAS DISPATCHED TO THE CUSTOMER SITE TO OBTAIN SCAN SPECIFIC INFORMATION AND INVESTIGATE THE ENVIRONMENTAL CONDITIONS. THE INVESTIGATION FOCUSED ON FOUR MAIN AREAS: ENVIRONMENTAL: (INCLUDING COOLING EQUIPMENT, PT AIR COOLING PLUS THE MR SCANNER ERROR LOG). SURFACE COILS/ACCESSORIES: (SURFACE COIL/ECG CHECKS). SYSTEM/ IMAGE QUALITY: (SYSTEMS LEVEL TESTING TO CHECK FOR SYSTEM FAILURES). PT MONITORING: (PT ALARM & RF SAFETY MONITOR CHECKS). VISUAL INSPECTION AND THE TEST RESULTS FOR THE MRI SYSTEM SHOW NO ISSUES THAT CAUSED OR CONTRIBUTED TO THE BURN. THE SYSTEM WAS DETERMINED TO BE FUNCTIONING WITHIN SPECIFICATIONS. THE ROOT CAUSE OF THE INJURY COULD NOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PT SUSTAINED A SECOND DEGREE BURN ON THE MIDDLE OF HER BACK TO THE RIGHT OF THE MIDLINE DURING A MAGNETIC RESONANCE (MR) SCAN. THE PT WAS BEING SCANNED WITH THE 8 CHANNEL CTL COIL FOR A LUMBAR SPINE SCAN. THE PT WAS POSITIONED HEAD FIRST AND SUPINE, AND THE PT'S ARMS WERE POSITIONED OVER HER HEAD. THE PT REPORTEDLY FELT REDNESS AND DISCOMFORT, BUT SHE DID NOT INDICATE THIS DIRECTLY AFTER HER MR SCAN. SHE CONTACTED THE SITE 10 DAYS AFTER THE SCAN. THE HOSPITAL STATED THAT THE PT WAS PROPERLY ISOLATED FROM THE BORE. THE COMBINED SIZE OF THE BLISTERS WAS 2 X 3 CM. THE PULSE SEQUENCES USED WERE NOT PROVIDED. THE PT DID NOT SEEK ANY MEDICAL TREATMENT FOR THIS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 1.5T SIGNA HDX ECHOSPEED MAGNETIC RESONANCE DIAGNOSTIC SYSTEM LNH GE MEDICAL SYSTEMS, LLC 5308573

Patients

Seq Age Sex Outcome Treatment
1 47 YR Other