CAPSURE SP
Report
- Report Number
- 2649622-2011-05575
- Event Type
- Injury
- Date Received
- April 12, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P850089/S15
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON SEVERAL CONDUCTORS (NOT OBSTRUCTED), THE INNER TUBING WAS KINKED/BUCKLED, THE OUTER INSULATION HAD A COSMETIC CUT, THERE WAS A WHITE SUBSTANCE ON THE EXPOSED DEFIBRILLATION COIL, THERE WAS A WHITE SUBSTANCE ON THE OUTER INSULATION, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THE HELIX/LOBE WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4): THE LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), THE OUTER INSULATION WAS PARTIALLY OR FULLY COVERING THE TIP ELECTRODE, THE OUTER INSULATION WAS BREACHED CUT, THERE WAS A WHITE SUBSTANCE ON THE OUTER INSULATION, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4): THE DISTAL PORTION OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), THE INNER INSULATION WAS PULLED APART DUE TO OVERSTRESS, THE OUTER INSULATION WAS PARTIALLY OR FULLY COVERING THE TIP ELECTRODE, THE OUTER INSULATION HAD A COSMETIC CUT, THE OUTER INSULATION WAS BREACHED CUT, THERE WAS A WHITE SUBSTANCE ON THE OUTER INSULATION, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS APPARENT EXPLANT DAMAGE AND THE LEAD WAS STRETCHED.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) HIGH VOLTAGE LEAD HAD HIGH THRESHOLDS. IT WAS FURTHER REPORTED THAT WHEN EXTRACTING THE RV HIGH VOLTAGE LEAD THE RIGHT VENTRICULAR AND RIGHT ATRIAL PACING LEADS WERE ADHERED TO IT AS WERE EXPLANTED AS WELL. ALL LEADS WERE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE SP | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5024M | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |