FDA Adverse Event Injury Summary report: N

SPRINT

MDR report key: 2051529 · Received April 12, 2011

Report

Report Number
2649622-2011-05573
Event Type
Injury
Date Received
April 12, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S13
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON SEVERAL CONDUCTORS (NOT OBSTRUCTED), THE INNER TUBING WAS KINKED/BUCKLED, THE OUTER INSULATION HAD A COSMETIC CUT, THERE WAS A WHITE SUBSTANCE ON THE EXPOSED DEFIBRILLATION COIL, THERE WAS A WHITE SUBSTANCE ON THE OUTER INSULATION, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THE HELIX/LOBE WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4): THE LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), THE OUTER INSULATION WAS PARTIALLY OR FULLY COVERING THE TIP ELECTRODE, THE OUTER INSULATION WAS BREACHED CUT, THERE WAS A WHITE SUBSTANCE ON THE OUTER INSULATION, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4): THE DISTAL PORTION OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), THE INNER INSULATION WAS PULLED APART DUE TO OVERSTRESS, THE OUTER INSULATION WAS PARTIALLY OR FULLY COVERING THE TIP ELECTRODE, THE OUTER INSULATION HAD A COSMETIC CUT, THE OUTER INSULATION WAS BREACHED CUT, THERE WAS A WHITE SUBSTANCE ON THE OUTER INSULATION, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS APPARENT EXPLANT DAMAGE AND THE LEAD WAS STRETCHED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) HIGH VOLTAGE LEAD HAD HIGH THRESHOLDS. IT WAS FURTHER REPORTED THAT WHEN EXTRACTING THE RV HIGH VOLTAGE LEAD THE RIGHT VENTRICULAR AND RIGHT ATRIAL PACING LEADS WERE ADHERED TO IT AS WERE EXPLANTED AS WELL. ALL LEADS WERE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6943 ASKU

Patients

Seq Age Sex Outcome Treatment
1 47 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB