SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-05569
- Event Type
- Injury
- Date Received
- April 12, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DISTAL CONDUCTOR WAS DISTORTED, THE DEFIBRILLATION CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON THE OUTER TUBING OVERLAY, THE INNER TUBING WAS KINKED/BUCKLED, THE OUTER TUBING OVERLAY WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THE HELIX/LOBE WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM, THERE WAS APPARENT EXPLANT DAMAGE AND THE LEAD WAS STRETCHED.
IT WAS REPORTED THE PATIENT RECEIVED INAPPROPRIATE SHOCKS DUE TO LEAD OVERSENSING WITH A LOW SENSING VALUE ON THE R WAVE. THE LEAD WAS REMOVED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR | Hospitalization| L| R | 7304 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4194 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD |