FDA Adverse Event Other Summary report: N

ASR ACETABULAR CUPS 48

MDR report key: 2051457 · Received March 9, 2011

Report

Report Number
1818910-2011-03741
Event Type
Other
Date Received
March 9, 2011
Date of Event
February 7, 2011
Report Date
February 7, 2011
Manufacturer
DEPUY INTERNATIONAL, LTD.
Product Code
KWA
PMA / PMN Number
K040627
Removal / Correction Number
Z-1749/1816-2011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS CUP LOOSENING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASR ACETABULAR CUPS 48 87KWA KWA DEPUY INTERNATIONAL, LTD. NA 2427277

Patients

Seq Age Sex Outcome Treatment
1 Female Required Intervention