FDA Adverse Event Injury Summary report: N

ATTAIN STARFIX

MDR report key: 2051421 · Received April 12, 2011

Report

Report Number
2649622-2011-05559
Event Type
Injury
Date Received
April 12, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P060039
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT ALL CONDUCTORS HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT, AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THE DISTAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED) AND THERE WAS BLOOD/BODY FLUID IN THE OUTER TUBING OVERLAY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD HAD HIGH THRESHOLD AND HAD NO CAPTURE. THE LEAD WAS REMOVED. IT WAS ALSO REPORTED THAT A NEW LEAD WAS ATTEMPTED, BUT IT WAS TOO TIGHT TO GO THROUGH THE VEIN. THE LEAD WAS NOT USED AND A DIFFERENT NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN STARFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4195 ASKU

Patients

Seq Age Sex Outcome Treatment
1 84 YR Hospitalization| O| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 1488TC IMPLANTABLE PACING LEAD| 7001 IMPLANTABLE DEFIB LEAD