FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM

MDR report key: 2051364 · Received April 12, 2011

Report

Report Number
3005075853-2011-01466
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
February 14, 2011
Report Date
February 16, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADVANCER. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FIRING OF THE DEVICE, THE TIP OF THE ADVANCER SLID TOWARD TISSUE STOP DURING FIVE CONSECUTIVE FIRING SEQUENCES CAUSING THAT THE JAWS REMAINED IN THE CLOSED. THE TRIGGER WAS MANUALLY ASSISTED IN ORDER TO OPEN THE JAWS WITHOUT ANY DIFFICULTIES NOTED; PEAR SHAPED CLIPS WERE RELEASED. THE NEXT THREE CLIPS WERE CONFORMING ACCORDING TO OUR MANUFACTURING SPECIFICATIONS. DURING THE LAST FIVE FIRING SEQUENCES SCISSOR CLIPS WERE FED. IN ORDER TO EVALUATE THE CONDITION OF THE INTERNAL COMPONENTS OF THE DEVICE, IT WAS DISASSEMBLED. UPON DISASSEMBLING, THE FEED CONNECTOR WELD WAS NOTED TO BE NOT PROPERLY ASSEMBLED CAUSING CLIP SHORT FED; THE JAWS GOT DISTORTED DURING THE FORMATION OF THE CLIPS CAUSING THE FOUND SCISSOR CLIPS. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE INDICATOR WAS NOT VISIBLE. THIS FINDING IS NOT RELATED WITH THE INCIDENT REPORTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE THE DEVICE LOCKED ON THE CYSTIC DUCT. THE DEVICE WAS REMOVED WITHOUT ANY DAMAGE TO THE DUCT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1