FDA Adverse Event
Malfunction
Summary report: N
HEARTSTART MRX
MDR report key: 2051344
·
Received April 4, 2011
Report
- Report Number
- 1218950-2011-00882
- Event Type
- Malfunction
- Date Received
- April 4, 2011
- Report Date
- March 9, 2011
- Manufacturer
- PHILIPS HEALTHCARE
- Product Code
- MKJ
- PMA / PMN Number
- K031187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE CUSTOMER REPORTED A CHARGE BUTTON FAILURE AND A BEEPING NOISE. THERE WAS NO REPORTED PT INVOLVEMENT. THE DEVICE WAS EVALUATED AND THE FIELD SVC ENGINEER VERIFIED THE CHARGE BUTTON FAILURE MESSAGE. THE ISSUE WAS RESOLVED BY RELOADING THE SOFTWARE AND REBOOTING THE DEVICE. WE ARE CONSIDERING THIS A MALFUNCTION RESULTING FROM CORRUPT SOFTWARE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A CHARGE BUTTON FAILURE AND A BEEPING NOISE. THERE WAS NO REPORTED PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSTART MRX | MKJ | PHILIPS HEALTHCARE | M3535A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |