FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE
Report
- Report Number
- 6000001-2011-02788
- Event Type
- Malfunction
- Date Received
- April 12, 2011
- Date of Event
- March 29, 2011
- Report Date
- March 29, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915523
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF THE COLLEAGUE INFUSION PUMP WITH A LOW BATTERY ALARM WAS CONFIRMED. THE ASSIGNABLE CAUSE WAS DEFECTIVE MAIN BATTERY. THE MAIN BATTERY WAS REPLACED TO CORRECT THE PROBLEM. ADDITIONAL: A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. (B)(4). BAXTER WILL CONTINUE TO MONITOR AND REPORT ON DEATH AND SERIOUS INJURY EVENTS AND FOLLOW EXISTING ESCALATION PROCESSES TO ASSESS THE IMPACT ON PATIENT SAFETY.
A BAXTER SERVICE TECHNICIAN REPORTED FINDING A FLO-GARD INFUSION PUMP WITH A LOW BATTERY ALARM OCCURRING DURING THE BATTERY OPERATION TEST. THIS ALARM INTERRUPTED DELIVERY. THIS EVENT OCCURRED DURING PRODUCT EVALUATION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |