FDA Adverse Event Malfunction Summary report: N

FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE

MDR report key: 2051280 · Received April 12, 2011

Report

Report Number
6000001-2011-02788
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
March 29, 2011
Report Date
March 29, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF THE COLLEAGUE INFUSION PUMP WITH A LOW BATTERY ALARM WAS CONFIRMED. THE ASSIGNABLE CAUSE WAS DEFECTIVE MAIN BATTERY. THE MAIN BATTERY WAS REPLACED TO CORRECT THE PROBLEM. ADDITIONAL: A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. (B)(4). BAXTER WILL CONTINUE TO MONITOR AND REPORT ON DEATH AND SERIOUS INJURY EVENTS AND FOLLOW EXISTING ESCALATION PROCESSES TO ASSESS THE IMPACT ON PATIENT SAFETY.

Description of Event or Problem · 1

A BAXTER SERVICE TECHNICIAN REPORTED FINDING A FLO-GARD INFUSION PUMP WITH A LOW BATTERY ALARM OCCURRING DURING THE BATTERY OPERATION TEST. THIS ALARM INTERRUPTED DELIVERY. THIS EVENT OCCURRED DURING PRODUCT EVALUATION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1