KAPPA 700 D
Report
- Report Number
- 6000094-2011-00488
- Event Type
- Injury
- Date Received
- April 12, 2011
- Date of Event
- January 12, 2011
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- DXY
- PMA / PMN Number
- P980035
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4): PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.
IT WAS REPORTED THAT THE BATTERY VOLTAGE OF THE DEVICE DEPLETED VERY QUICKLY TOWARDS THE END OF THE BATTERY LIFE. THE DEVICE COULD NOT BE INTERROGATED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | KAPPA 700 D | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC S.A. | KD701 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | Hospitalization| R |