FDA Adverse Event Injury Summary report: N

KAPPA 700 D

MDR report key: 2051273 · Received April 12, 2011

Report

Report Number
6000094-2011-00488
Event Type
Injury
Date Received
April 12, 2011
Date of Event
January 12, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4): PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BATTERY VOLTAGE OF THE DEVICE DEPLETED VERY QUICKLY TOWARDS THE END OF THE BATTERY LIFE. THE DEVICE COULD NOT BE INTERROGATED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KAPPA 700 D IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. KD701 ASKU

Patients

Seq Age Sex Outcome Treatment
1 84 YR Hospitalization| R