TARGET COIL
Report
- Report Number
- 2939204-2011-00213
- Event Type
- Injury
- Date Received
- April 12, 2011
- Date of Event
- March 31, 2011
- Report Date
- March 31, 2011
- Manufacturer
- BOSTON SCIENTIFIC - CORK
- Product Code
- HCG
- PMA / PMN Number
- K093142
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
A DEVICE HISTORY RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. ONLY THE DELIVERY WIRE WAS RETURNED FOR ANALYSIS AND THIS WAS RETURNED WITH THE MICROCATHETER USED IN THE PROCEDURE, NO ANOMALIES WERE NOTED TO THE MICROCATHETER. THE MAIN COIL REMAINS IMPLANTED. VISUAL INSPECTION OF THE DEVICE NOTED THAT THE DEVICE WAS EXTENSIVELY STRETCHED ON THE INNER CLOSED PITCH COIL, DISTAL TO THE DETACHMENT ZONE. THE DETACHMENT ZONE WAS FRACTURED. THE POLYIMIDE (PI) COATING ON THE DELIVERY WIRE WAS DETACHED AS A RESULT OF THE EXCESSIVE STRETCHING OF THE INNER COIL. THE OUTER OPEN PITCH COIL AND PLATINUM MARKER COIL WERE INTACT AND UNDAMAGED. APPROXIMATELY 25CM OF THE INNER COIL, PROXIMAL TO THE PLATINUM MARKER COIL WAS ALSO INTACT. THE DELIVERY WIRE WAS KINKED IN SEVERAL LOCATIONS ALONG ITS LENGTH. EXAMINATION UNDER MAGNIFICATION CONFIRMED THE INNER CLOSED PITCH COIL DISTAL TO THE DETACHMENT ZONE WAS STRETCHED AND DETACHED FROM THE DELIVERY WIRE. FROM EXAMINATION OF THE DISTAL END OF THE DELIVERY WIRE IT WAS EVIDENT THAT THE MAIN COIL HAD DETACHED FROM THE DELIVERY WIRE AT THE DETACHMENT ZONE. THE PI COATING ON THE WIRE WAS NOTED TO BE EXTENDING FROM THE PEBAX AND TO BE BROKEN. THE FIRM COIL WAS EXTENSIVELY STRETCHED AND UNRAVELED FROM THE DETACHMENT ZONE. BASED ON THE INFORMATION PROVIDED AND THE INVESTIGATION IT IS MOST LIKELY THAT THE DELIVERY WIRE STRETCHED AND BROKE AFTER RESISTANCE WAS ENCOUNTERED IN THE MICROCATHETER AS IT WAS BEING REMOVED FROM THE PATIENT. IT WAS CONCLUDED THAT THE MOST PROBABLE CAUSE FOR THE REPORTED EVENT AND OBSERVED DAMAGE TO THE DEVICE WAS OPERATIONAL CONTEXT.
DURING THE PROCEDURE, THE COIL BROKE FROM THE PUSHER WIRE AND A LOOP OF THE COIL PROTRUDED INTO THE PARENT VESSEL. THE PHYSICIAN ATTEMPTED TO SECURE THE LOOP TO THE PARENT VESSEL WALL BUT THIS WAS NOT SUCCESSFUL. THE PROCEDURE WAS COMPLETED BY COILING THE ANEURYSM AND THE COIL LOOP REMAINED PROTRUDING INTO THE PARENT VESSEL. THE PATIENT WAS REPORTED TO BE NEUROLOGICALLY INTACT.
DURING THE PROCEDURE, THE COIL BROKE FROM THE PUSHER WIRE AND A LOOP OF THE COIL PROTRUDED INTO THE PARENT VESSEL. THE PHYSICIAN ATTEMPTED TO SECURE THE LOOP TO THE PARENT VESSEL WALL BUT THIS WAS NOT SUCCESSFUL. THE PROCEDURE WAS COMPLETED BY COILING THE ANEURYSM AND THE COIL LOOP REMAINED PROTRUDING INTO THE PARENT VESSEL. THE PATIENT WAS REPORTED TO BE NEUROLOGICALLY INTACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TARGET COIL | DEVICE, NEUROVASCULAR EMBOLIZATION | HCG | BOSTON SCIENTIFIC - CORK | M0035461430 | 14044772 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Required Intervention |