FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX

MDR report key: 2051032 · Received April 12, 2011

Report

Report Number
2649622-2011-05472
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
January 26, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S2
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR WAS DISTORTED. IT WAS NOTED THAT THE LEAD WAS RECEIVED WITH THE HELIX FULLY RETRACTED. WHEN THE HELIX WAS RETRACTED DURING THE PROCEDURE, THE IS-1 CONNECTOR PIN WAS TURNED AN EXCESSIVE NUMBER OF TIMES, RESULTING IN DISTORTION OF THE DISTAL CONDUCTOR WITHIN THE CONNECTOR. IT WAS ALSO NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD APPEARED DAMAGED AT IMPLANT. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR WAS DISTORTED. IT WAS NOTED THAT THE LEAD WAS RECEIVED WITH THE HELIX FULLY RETRACTED. WHEN THE HELIX WAS RETRACTED DURING THE PROCEDURE, THE IS-1 CONNECTOR PIN WAS TURNED AN EXCESSIVE NUMBER OF TIMES, RESULTING IN DISTORTION OF THE DISTAL CONDUCTOR WITHIN THE CONNECTOR. IT WAS ALSO NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD APPEARED DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE ATRIAL LEAD SHOWED HIGH IMPEDANCE AND WAS REPOSITIONED SEVERAL TIMES. THE HELIX WOULD NOT DEPLOY AND THE HELIX FIXATION MECHANISM FROZE. THE LEAD WAS NOT IMPLANTED AND AN ATTEMPT WAS MADE TO PLACE A NEW LEAD, BUT THE SECOND LEAD ALSO HAD THE SAME ISSUES. A THIRD LEAD WAS IMPLANTED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5568 ASKU

Patients

Seq Age Sex Outcome Treatment
1 85 YR Other