NT821707, DRAINAGE KIT, LUMBAR
Report
- Report Number
- 2023988-2024-00073
- Event Type
- Malfunction
- Date Received
- October 22, 2024
- Date of Event
- August 1, 2024
- Report Date
- November 27, 2024
- Manufacturer
- NATUS MEDICAL INCORPORATED
- Product Code
- JXG
- UDI-DI
- 00382830050531
- PMA / PMN Number
- K964923
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
FOLLOW UP REPORT 001 REF TO NATUS COMPLAINT#: (B)(4). DEVICE STERILE DATE: ON (B)(6) 2024. COMPLAINT HISTORY REVIEW/TREND ANALYSIS: (24 MONTHS REVIEW AND PERCENT OF SALES) 2 PREVIOUSLY CONFIRMED "INTEG-BROKE IN USE" COMPLAINTS WITHIN THE PAST TWO YEARS. (B)(4) NT821707 UNITS SOLD WITHIN PAST TWO YEARS. FAILURE RATE: (B)(4). CAPA: 005401 IS NOTED TO BE RELATED, CAPA WAS CLOSED IN 2022. DEVICE HISTORY RECORD REVIEW: UNIT HAS PASSED ALL THE REQUIRED TESTS. NO MANUFACTURING DEFECTS OR NON-CONFORMANCE'S OBSERVED. ROOT CAUSE/FAILURE INVESTIGATION: THE CUSTOMER DID NOT RETURN THE DEVICE FOR EVALUATION OR PROVIDE ANY FURTHER INFORMATION REGARDING THE ALLEGED COMPLAINT. FAILURE MODE: NO DEVICE RETURNED - UNABLE TO DETERMINE.
INITIAL REPORT REF TO NATUS COMPLAINT (B)(4). DEVICE IDENTIFIER (UDI NUMBER) : N/A. RISK REVIEW: (B)(4). REV 05 NATUS EXTERNAL DRAINAGE LUMBAR CATHETERS - RISK ANALYSIS SPREADSHEET HAZARD 1.5 - CATHETER TUBING IS CUT BY SHARP INNER EDGE OF NEEDLE WHILE ACCESSING INTO LUMBAR SPINE AND A CATHETER PIECE IS LEFT IN THE PATIENT BODY. EFFECT (HARM): SURGICAL INTERVENTION. RESIDUAL RISK: MEDIUM. FURTHER INVESTIGATION TO BE CARRIED OUT.
NT821707, DRAINAGE KIT, LUMBAR WHILE PERFORMING A LUMBAR DRAIN AT THE BEDSIDE - A PIECE OF THE SOFT TUBING WAS SHEARED AND LEFT IN THE THECAL SAC. THE DEVICE DID NOT DO WHAT IT WAS SUPPOSED TO DO; DEVICE WAS HARD TO USE.
PART 821707 CODMAN* LUMBAR CATHETER KIT - WHILE PERFORMING A LUMBAR DRAIN AT THE BEDSIDE - A PIECE OF THE SOFT TUBING WAS SHEARED AND LEFT IN THE THECAL SAC. THE DEVICE DID NOT DO WHAT IT WAS SUPPOSED TO DO; DEVICE WAS HARD TO USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1790573 | NT821707, DRAINAGE KIT, LUMBAR | NT821707, DRAINAGE KIT, LUMBAR | JXG | NATUS MEDICAL INCORPORATED | NT821707 | 118001039913 | 00382830050531 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Female |