PROXIMATE** RELOADABLE LINEAR STAPLER - 60 MM
Report
- Report Number
- 3005075853-2011-01463
- Event Type
- Malfunction
- Date Received
- April 12, 2011
- Date of Event
- February 14, 2011
- Report Date
- March 21, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD PRESENT. THE RELOAD WAS RECEIVED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLES WERE NOTED TO HAVE A PROPER B-FORMATION AND THE STAPLE LINE WAS COMPLETE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN (B)(4), AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A HARTMANN PROCEDURE, THE DEVICE WAS FIRED AND THE STAPLE LINE WAS NOT CLOSED. NO INFORMATION ON THE STAPLE FORMATION. THE SURGEON USED SUTURE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT REPORTED. NO OTHER DETAILS OF THE EVENT ARE KNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROXIMATE** RELOADABLE LINEAR STAPLER - 60 MM | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | G4UA1X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARTRIDGE |