FDA Adverse Event Malfunction Summary report: N

PROXIMATE** RELOADABLE LINEAR STAPLER - 60 MM

MDR report key: 2050938 · Received April 12, 2011

Report

Report Number
3005075853-2011-01463
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
February 14, 2011
Report Date
March 21, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD PRESENT. THE RELOAD WAS RECEIVED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLES WERE NOTED TO HAVE A PROPER B-FORMATION AND THE STAPLE LINE WAS COMPLETE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN (B)(4), AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A HARTMANN PROCEDURE, THE DEVICE WAS FIRED AND THE STAPLE LINE WAS NOT CLOSED. NO INFORMATION ON THE STAPLE FORMATION. THE SURGEON USED SUTURE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT REPORTED. NO OTHER DETAILS OF THE EVENT ARE KNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE** RELOADABLE LINEAR STAPLER - 60 MM STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA G4UA1X

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE