FDA Adverse Event Malfunction Summary report: N

ENDOPATH** XCEL* TROCAR

MDR report key: 2050910 · Received April 12, 2011

Report

Report Number
3005075853-2011-01460
Event Type
Malfunction
Date Received
April 12, 2011
Report Date
March 22, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: WERE ANY NOISES HEARD SUCH AS "WHISTLING" OR "HISSING"? IF SO, DID THE "NOISE" PREVENT INSUFFLATION? PLEASE DESCRIBE THE NOISE. ASKU. WAS THERE A DROP IN PRESSURE? IF YES, DID THIS AFFECT THE VISIBILITY OF THE SURGEON? ASKU. WHAT WAS THE GAS CONSUMPTION RATE OR VOLUME (LITER/MINUTE)? ASKU. WERE YOU ABLE TO IDENTIFY WHERE THE LEAK WAS COMING FROM? ASKU. WAS A DEVICE INSERTED IN THE TROCAR DURING THE LEAKING? IF SO, WHAT DEVICE? ASKU. WAS ANY TORQUE BEING APPLIED TO THE TROCAR OR DEVICE? ASKU. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS FUNCTIONALLY TESTED TO DETECT ANY LEAKING ISSUES. UPON EVALUATION OF THE DEVICE, IT WAS FUNCTIONALLY LEAK TESTED AND PASSED. THE DEVICE WAS FULLY FUNCTIONAL ACCORDING TO THE MANUFACTURING REQUIREMENTS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE CAUSED INSUFFLATION ISSUES. NO OTHER INFORMATION IS AVAILABLE. IT IS UNKNOWN HOW THE CASE WAS COMPLETED. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** XCEL* TROCAR LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1