FDA Adverse Event Injury Summary report: N

ZIMMON PANCREATIC STENT

MDR report key: 2050899 · Received April 1, 2011

Report

Report Number
3001845648-2011-00013
Event Type
Injury
Date Received
April 1, 2011
Report Date
March 7, 2011
Manufacturer
COOK IRELAND LTD
Product Code
FGE
PMA / PMN Number
K900923
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

METHOD: THE DEVICE INVOLVED IN THIS COMPLAINT IS AVAILABLE FOR EVALUATION BUT TO-DATE, IT HAS NOT BEEN REC'D AT COOK (B)(4) (THE MANUFACTURER). A DETAILED EVALUATION WILL BE CARRIED OUT WHEN THE DEVICE IS REC'D AT THE MANUFACTURER SITE. THE DEVICE INVOLVED IN THIS COMPLAINT IS AN (B)(4) (ZIMMON PANCREATIC STENT) DEVICE. THE ACTUAL LOT NUMBER OF THE DEVICE INVOLVED IN THIS COMPLAINT WAS NOT PROVIDED. AS THE LOT NUMBER WAS NOT PROVIDED; IT WAS THEREFORE NOT POSSIBLE TO ESTABLISH IF THERE WERE ANY DEVICES FROM THE AFFECTED LOT NUMBER IN STOCK AT THE TIME OF THE COMPLAINT INVESTIGATION. WE WERE INFORMED THAT THE DEVICE INVOLVED IN THIS COMPLAINT WAS AVAILABLE TO BE RETURNED FOR EVALUATION; HOWEVER, THE DEVICE WAS NOT REC'D AT COOK (B)(4) TO-DATE. AS THE DEVICE WAS NOT REC'D, THE CUSTOMER COMPLAINT COULD NOT BE CONFIRMED AND THE CAUSE OF THE COMPLAINT COULD NOT BE CONCLUSIVELY DETERMINED. WITH THE INFORMATION PROVIDED A DOCUMENT BASED INVESTIGATION WAS CARRIED OUT. PRIOR TO DISTRIBUTION, ZIMMON PANCREATIC STENT ARE SUBJECTED TO VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE MANUFACTURING RECORDS FOR THE (B)(4) DEVICE INVOLVED IN THIS COMPLAINT COULD NOT BE REVIEWED AS THE LOT NUMBER COULD NOT BE PROVIDED. NO CORRECTIVE ACTION IS WARRANTED AT THIS TIME. A DEFINITIVE CAUSE FOR THIS OBSERVATION WAS UNABLE TO BE DETERMINED AS THE DEVICE HAS NOT YET BEEN REC'D FOR EVALUATION. DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PT ANATOMY OR PROGRESSION OF DISEASE STATE, COOK (B)(4) COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF DEVICE USAGE, HOWEVER, CUSTOMER QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.

Description of Event or Problem · 1

WHILE TRYING TO REMOVE A ZIMMON PANCREATIC STENT (5F 7CM SINGLE PIGTAIL STENT), THE STENT SEPARATED IN THE PT. THE PIGTAIL WAS CLEARLY OUTSIDE IN THE DUODENUM. THE TIP OF THE STENT WAS GRASPED WITH A SNARE AND THE STENT WAS PULLED. THE STENT FRACTURED SOMEWHERE CLOSE TO THE START OF THE PIGTAIL/RADIOOPAQUE MARKER. THE STENT DID NOT FRACTURE WHERE THE STENT WAS GRASPED WITH THE SNARE. ADDITIONAL PHYSICIANS WERE CALLED IN TO ASSIST; AFTER TIME THEY WERE ABLE TO REMOVE THE STENT WITH FORCEPS. THE PT IS FINE; HOWEVER, THE PT DID EXPERIENCE ADDITIONAL STRESS DUE TO THE EXTENDED PERIOD OF TIME REQUIRED IN THE PROCEDURE. THE PT WAS DISCHARGED IN GOOD CONDITION, FULL RECOVERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ZIMMON PANCREATIC STENT FGE CATHETER, BILIARY, DIAGNOSTIC FGE COOK IRELAND LTD UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention