TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO
Report
- Report Number
- 2953200-2011-00738
- Event Type
- Injury
- Date Received
- April 1, 2011
- Date of Event
- March 4, 2011
- Report Date
- March 4, 2011
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). RESULTS: (STENT GRAFT MIGRATION, ENDOLEAK). RESULTS & CONCLUSIONS: DISEASE PROGRESSION AND AORTIC NECK DILATATION).
A TALENT STENT GRAFT WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROXIMATELY 15 MONTHS AGO. ANEURYSM MORPHOLOGY AT THE TIME OF IMPLANT IS UNK. THE AORTIC NECK IS REVERSE TAPERED SHAPED RANGING FROM 22 MM - 27 MM AND HAS GOTTEN WORSE INCREASING 1 MM IN DIAMETER EVERY 3 MM SLICE OF THE CT SCAN. IT WAS REPORTED THAT RECENT FILMS SHOW THE TALENT BIFURCATED SENT GRAFT IS 3 CM BELOW THE LEVEL OF THE RENAL ARTERIES AND THERE IS A TYPE I ENDOLEAK PRESENT. THE STENT GRAFT MIGRATION WAS ATTRIBUTED TO DISEASE PROCESSING AND AORTIC NECK DILATATION. A 70 MM ENDURANT CUFF WAS PLACED AND SUCCESSFULLY RESOLVED THE STENT GRAFT MIGRATION/ENDOLEAK. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO | MIH | MEDTRONIC IRELAND | NA | V00388435 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Required Intervention |