FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO

MDR report key: 2050862 · Received April 1, 2011

Report

Report Number
2953200-2011-00738
Event Type
Injury
Date Received
April 1, 2011
Date of Event
March 4, 2011
Report Date
March 4, 2011
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: (STENT GRAFT MIGRATION, ENDOLEAK). RESULTS & CONCLUSIONS: DISEASE PROGRESSION AND AORTIC NECK DILATATION).

Description of Event or Problem · 1

A TALENT STENT GRAFT WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROXIMATELY 15 MONTHS AGO. ANEURYSM MORPHOLOGY AT THE TIME OF IMPLANT IS UNK. THE AORTIC NECK IS REVERSE TAPERED SHAPED RANGING FROM 22 MM - 27 MM AND HAS GOTTEN WORSE INCREASING 1 MM IN DIAMETER EVERY 3 MM SLICE OF THE CT SCAN. IT WAS REPORTED THAT RECENT FILMS SHOW THE TALENT BIFURCATED SENT GRAFT IS 3 CM BELOW THE LEVEL OF THE RENAL ARTERIES AND THERE IS A TYPE I ENDOLEAK PRESENT. THE STENT GRAFT MIGRATION WAS ATTRIBUTED TO DISEASE PROCESSING AND AORTIC NECK DILATATION. A 70 MM ENDURANT CUFF WAS PLACED AND SUCCESSFULLY RESOLVED THE STENT GRAFT MIGRATION/ENDOLEAK. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC IRELAND NA V00388435

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention