FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 2050852 · Received April 4, 2011

Report

Report Number
1824206-2011-01996
Event Type
Malfunction
Date Received
April 4, 2011
Date of Event
March 10, 2011
Report Date
March 10, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THAT FLUID WAS SPLASHED ONTO THE CAREGIVER CONTROLS ON THE LEFT SIDE RAIL. THE TECHNICIAN REPLACED THE LEFT SIDE RAIL POWER CONTROL BOARD TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

TECHNICIAN ALLEGED THAT THE LEFT CAREGIVER SIDE RAIL HAD NO FUNCTION. NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1