FDA Adverse Event Injury Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL

MDR report key: 2050789 · Received April 1, 2011

Report

Report Number
2023826-2011-00280
Event Type
Injury
Date Received
April 1, 2011
Date of Event
March 7, 2011
Report Date
March 8, 2011
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P880091
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION METHOD: WORK ORDER SEARCH. RESULTS: A LENS WORK ORDER SEARCH WAS PERFORMED AND THERE WERE NO SIMILAR COMPLAINTS FOUND. CONCLUSION: NO CONCLUSION CAN BE DRAWN. BASED ON THE COMPLAINT HISTORY AND WORK ORDER SEARCH, WE WERE UNABLE TO DETERMINE A SPECIFIC ROOT CAUSE OF THE EVENT. (B)(4).

Additional Manufacturer Narrative · 1

THE SCRUB TECH ON THIS CASE CALLED BACK AND STATED THE EYE WAS IN GOOD CONDITION PRIOR SURGERY AND WHEN THE LENS WAS INSERTED THE CAPSULE TORE AND THE LENS COLLAPSED IN THE EYE. THE SURGEON RETRIEVED THE LENS AND PERFORMED A VITRECTOMY. AN ACL WAS IMPLANTED. THE CAUSE OF THE EVENT REMAINS UNKNOWN. EVALUATION RESULTS: VISUAL INSPECTION OF THE RETURNED LENS SHOWED EVIDENCE OF A CLEAR SURGICAL RESIDUE, HOWEVER, THERE WAS NO VISIBLE DAMAGE TO THE LENS. BOTH SIDES OF THE OPTIC/ HAPTIC WERE CHECKED FOR ROUGH/SHARP EDGES AND EDGES WERE FOUND TO BE ACCEPTABLE.

Description of Event or Problem · 1

THE REPORTER STATED THE AA4203TL SILICONE SINGLE PIECE LENS WAS IMPLANTED AND THE CAPSULE BROKE. FELL BACK AND THE IOL HAD TO BE REMOVED. A VITRECTOMY WAS NEEDED. ANOTHER LENS WAS IMPLANTED INTO THE EYE. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT YET RECEIVED. IF ADDITIONAL INFORMATION IS OBTAINED A SUPPLEMENTAL MEDWATCH FORM WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY AA4203TL NA

Patients

Seq Age Sex Outcome Treatment
1 54 YR Female Required Intervention CARTRIDGE: MODEL UNKNOWN, LOT # UNKNOWN| INJECTOR: MODEL UNKNOWN, LOT # UNKNOWN| INJECTOR: MODEL UNKNOWN, LOT # UNKNOWN| CARTRIDGE: MODEL UNKNOWN, LOT # UNKNOWN