FDA Adverse Event Malfunction Summary report: N

SENTINOL NITINOL BILIARY STENT SYSTEM

MDR report key: 2050665 · Received April 12, 2011

Report

Report Number
2134265-2011-01230
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
March 17, 2011
Report Date
March 18, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
FGE
PMA / PMN Number
K032025
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4)

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: RETURNED PRODUCT CONSISTED OF A SENTINOL STENT DELIVERY SYSTEM WITH NO ORIGINAL PACKAGING AND A GUIDEWIRE FROZE INSIDE THE SENTINOL DEVICE. VISUAL AND MICROSCOPIC ANALYSIS REVEALED BLOOD WAS VISIBLE INSIDE THE INNER HUB AND THROUGHOUT THE ENTIRE LENGTH OF THE DEVICE. THE SENTINOL DEVICE WAS PLACED IN A WARM CIRCULATING WATER BATH FOR 48 HOURS TO TRY AND SEPARATE THE DEVICES. THIS WAS UNSUCCESSFUL, THE GUIDEWIRE WAS NOT ABLE TO BE REMOVED FROM THE SENTINOL DEVICE. APPROXIMATELY 80.5CM OF THE GUIDEWIRE WAS MEASURED PROTRUDING FROM THE PROXIMAL HUB AND 22CM PROTRUDING FROM THE TIP. THERE IS CLEARANCE BETWEEN THE OD OF THE GUIDEWIRE AND THE ID OF THE INNER MEMBER, AS A LAYER OF BLOOD FILLS THIS GAP CAUSING THE WIRE TO STICK. THIS DELIVERY SYSTEM IS COMPATIBLE WITH A 0.035 INCH GUIDEWIRE, WITH THE ACTUAL GUIDEWIRE TIED TO THIS COMPLAINT MEASURING APPROXIMATELY 0.035 AT ITS MAXIMUM. THERE ARE NO INDICATIONS OF NONCONFORMANCE AS PERTAINING TO THE SENTINOL STENT DELIVERY SYSTEM, OR THE GUIDEWIRE, AND IT APPEARS THAT BLOOD LODGED BETWEEN THE INNER-MEMBER ID AND THE GUIDEWIRE OD MAY BE THE CAUSING FACTOR OF THE WIRE BEING FROZE INSIDE THE SENTINOL DEVICE. THE STENT WAS RETURNED DEPLOYED AND RETURNED WITH THE SENTINOL DEVICE. MICROSCOPIC INSPECTION REVEALED NO DAMAGE OR IRREGULARITIES TO THE STENT OR THE DISTAL END OF THE SHAFT. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE WAS DETERMINED TO BE USER RELATED AS THE SENTINOL BILIARY PRODUCT IS INDICATED ONLY FOR USE IN BILIARY STENTING PROCEDURES AND THE RETURNED DEVICE WAS REPORTEDLY USED IN A PROCEDURE TO TREAT A SUPERFICIAL FEMORAL ARTERY. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE THE STENT DELIVERY SYSTEM BECAME STUCK ON THE GUIDE WIRE. THE 100% STENOSED DENOVO LESION WAS LOCATED IN THE OSTIUM OF THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA). WHILE LOADING THE 7X20X1350 SENTINOL STENT DELIVERY SYSTEM ONTO A NON BSC GUIDE WIRE, THE DEVICE BECAME STUCK AT ABOUT 100CM ON THE WIRE. UNABLE TO ADVANCE OR WITHDRAW THE STENT DELIVERY SYSTEM BOTH DEVICES WERE REMOVED TOGETHER AS ONE UNIT. ANOTHER NON BSC GUIDE WIRE AND STENT WERE USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS FINE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE THE STENT DELIVERY SYSTEM BECAME STUCK ON THE GUIDE WIRE. THE 100% STENOSED DENOVO LESION WAS LOCATED IN THE OSTIUM OF THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA). WHILE LOADING THE 7X20X1350 SENTINOL STENT DELIVERY SYSTEM ONTO A NON BSC GUIDE WIRE, THE DEVICE BECAME STUCK AT ABOUT 100CM ON THE WIRE. UNABLE TO ADVANCE OR WITHDRAW THE STENT DELIVERY SYSTEM BOTH DEVICES WERE REMOVED TOGETHER AS ONE UNIT. ANOTHER NON BSC GUIDE WIRE AND STENT WERE USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SENTINOL NITINOL BILIARY STENT SYSTEM CATHETER, BILIARY, DIAGNOSTIC FGE BOSTON SCIENTIFIC - MAPLE GROVE H7493894872010 12854438

Patients

Seq Age Sex Outcome Treatment
1 63 YR GUIDE WIRE, VERSA .035| INTRODUCER SHEATH, BALKIN 7F