FDA Adverse Event Malfunction Summary report: N

PEN NDL 32G 4MM PRO 90CT US

MDR report key: 20506208 · Received October 22, 2024

Report

Report Number
3023359743-2024-00586
Event Type
Malfunction
Date Received
October 22, 2024
Report Date
January 14, 2025
Manufacturer
EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND
Product Code
FMI
UDI-DI
00382903205745
PMA / PMN Number
K212015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

CORRECTION TO: H6 (COMPONENT CODE, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS) INVESTIGATION SUMMARY: SAMPLES WERE RECEIVED AND AN INVESTIGATION WAS PERFORMED. THIS IS THE 3RD COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE INDICATED ISSUE AS BENT CANNULA NPE. BASED ON TREND ANALYSIS NO FURTHER ACTION IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED.

Description of Event or Problem · 0

CONSUMER REPORTED NEEDLE CLOG DURING INJECTION. CONSUMER DOES NOT PRIME, HE DOES NOT RE-USE. LOT #: 3052557; CATALOG #: 320574; DATE OF EVENT: UNKNOWN; SAMPLES: AVAILABLE - SENDING MAIL KIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1957986 PEN NDL 32G 4MM PRO 90CT US NEEDLE, HYPODERMIC, SINGLE LUMEN FMI EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND 320574 3052557 00382903205745

Patients

Seq Age Sex Outcome Treatment
1 NA Male